Financial Incentives for 蜜桃传媒破解版下载 Labs

Our incentive program is for the purchase of energy or water-efficient equipment. If a lab plans to purchase or replace equipment, the lab can contact us and apply for a financial incentive toward the new equipment. This conversation needs to happen before the new purchase takes place. The value of the financial incentive is equivalent to up to five years of energy or water savings that 蜜桃传媒破解版下载 will gain from purchasing more efficient equipment. The incentive applies at the time of the equipment purchase. Incentives can take time and research to process. Please give us as much lead time as possible before your purchase.

Water Savings
  • Purchasing water efficient equipment
  • Replacing the use of a faucet aspirators with a vacuum pump
  • Replacing water-cooled condensers with waterless condensers
  • Free low flow aerators for lab faucets
Energy Savings
  • Purchasing energy efficient lab equipment
  • Covering disposal costs for lab equipment being retired without replacement

Examples of incentives given to 蜜桃传媒破解版下载 labs

The following are examples of incentives provided by CU Green Labs through funding provided by Facilities Management to laboratories at 蜜桃传媒破解版下载. Incentives may be available for the purchase of more energy or water efficient equipment or for the implementation of practices that allow for greater energy or water conservation. If you are interested in whether a financial incentive might be available to your lab, please reach out to听greenlabs@colorado.edu. 听Incentives can take time and research to process. Please give us as much lead time as possible prior to your purchase.听


WATER SAVING EXAMPLES:

罢丑别听Vaida Lab听in the听颁丑别尘颈蝉迟谤测听was able to get an oil-free membrane vacuum pump to replace their water aspirator for a savings of about 66,000 gallons of water per year.

A free Findenser was provided to the听Weber Lab听补迟听JILA听to be used instead of a water condenser. For a faucet flowing at a rate of .66 gallons per minute, a normal water condenser would use 158 gallons of water over four hours. A Findenser uses no water to cool chemical reactions for significant water savings.

ENERGY SAVING EXAMPLES:

A $3,670 incentive was provided to the听Fleshner Lab听颈苍听Integrative Physiology听towards a Stirling Ultra Low freezer to replace an inefficient ULT freezer. Another Stirling ULT freezer set at 鈥70 鈦癈听was metered as using 6.3 kWh of electricity per day whereas the old unit used 25.4 kWh per day.

A $400 incentive was provided to the听Fierer Lab听in the听Ecology and Evolutionary Biology Department听to replace their Fisher Scientific Isotemp Lab Refrigerator that used 5.9 kWh per day with an Energy Star unit that uses ~2 kWh per day.

A $1,075 incentive was provided for the听Neu Lab听颈苍听Mechanical Engineering听to purchase an energy efficient Biosafety Cabinet as their lab was being established. The Thermo Class II A2 BSC purchased by the Neu Lab has several energy efficient features. The Neu Lab is committed to turn off the unit at night and on the weekends and keeping the sash lowered between uses.

A $3,252 incentive was provided to the听McKnight Lab听补迟听INSTAAR听towards a cold temperature environmental chamber that cost $7,419. The Lunaire unit the lab had in place metered at 33.2 kWh per day whereas the replacement Powers Scientific DS52SD uses ~11.2 kWh per day for significant energy savings.

A free 鈥20 鈦癈听Energy Star Freezer was provided to the听Johnson Lab听at the听Institute for Behavioral Genetics听to replace an ULT freezer. Traditional ULT freezers use ~15-23 kWh of electricity per day whereas a -20 鈦癈听Energy Star freezer uses ~2 kWh per day.

CU Green Labs provided $266 towards a $799 Heidolph Brinkmann B-169 recirculating aspirator for the听Wuttke Lab in Biochemistry. A savant chiller consumes 5.5 kWh/day when left on all the time and a vacuum pump consumes 4-5 kWh/day for a total consumption of 10 kWh/day for those pieces of equipment combined. The recirculating aspirator is much easier to turn off when not in use, but would consume 5.5 kWh/day if left on all the time. The savings of using a recirculating aspirator is much more than 4.5 kWh/day since it can be turned off and on more easily. 5 years of 10 kWh/day (with Savant & vacuum pump) is 18250 kWh and 5 years of use of Heidolph Brinkmann B-169 recirculating aspirator for 1 hour per day, 5 days per week is 299 kWh. This is a difference of 17591 kWh over 5 years or $1,454 in electricity savings.

CU Green Labs supported sustainable research in the听Jimenez Lab听in the Department of Chemistry & Biochemistry听by providing a $1000 incentive to purchase a Thermo Scientific Heratherm drying oven. The new oven uses 5.04 kWh/day whereas other drying ovens metered on campus for energy consumption consume an average of听8.75 kWh/day. Not only is this drying oven more energy efficient than other models on the market,听the Jimenez group is further committed to energy conservation by turning off the oven overnight and during weekends. The oven comes with a timer that does this automatically.